Examples:
Data Sent Data Returned Meaning
W172<CR> W172<CR><LF> Echo of Command Sent
(Set Pressure units to “inH
2
O”)
W1700002<CR> W1700002<CR><LF> Echo of Command Sent
(Set Pressure units to inH
2
O”)
W01100<CR> W01100<CR><LF> Echo of Command Sent
??<CR><LF> Invalid Command Response
(Location 01 is Read Only)
U - UPLOAD
The UPLOAD command allows the user to read all of the selected
device data from locations 1 through 16 with a single command. The
data returned will be formatted as a single block per location
separated by a carriage return, line feed character sequence (CRLF
equivalent to hexadecimal 0x0D0A). See the table below for details
on the data structure.
Usage:
U(CR)
Where:
U UPLOAD command
CR - Carriage Return
Q - QUICKSEND
QUICKSEND is a feature that allows the user to receive an
automatic update of all of the meter data without any further user
interaction. When the QUICKSEND command is received, the
feature is turned ON and the meter will automatically send all of the
device data every half second. The Quicksend feature is toggled ON
and OFF with the QUICKSEND command. See the table below for
details on the data structure.
Usage:
Q(CR)
Where:
Q QUICKSEND command
CR - Carriage Return
V - VERSION
The VERSION command allows the user to read the Software Version
that the unit is currently running.
Usage:
V(CR)
Where:
V VERSION command
CR - Carriage Return