2-32
but only one datapoint would be stored every 10 seconds.
(See example report on next page.)
When using the Manual Datalog interval with the Nurse Call
feature of the Distal Occlusion test, the system will store three
datapoints to the Datalog after the occurrence of the Nurse
Call input. Therefore, the Distal Occlusion test will not
automatically stop for three times the duration of the LOG DATA
EVERY parameter. This is done to verify that the pressure
drops, indicating the pump has turned off, after the Nurse Call
input.
PAGE 01
BC GROUP
IPA-2000S INFUSION PUMP ANALYZER
PUMP INFORMATION:
MANUFACTURER: Manufacturer 1 MODEL: Pump 1
SERIAL NUMBER: 1001
DEPARTMENT: SURGERY LOCATION: Hospital 1
TEST STARTED AT: 7:13:26 ON 7/17/02
CHANNEL 1 FLOW RATE TEST
ELAPSED FLOW AVG FLOW VOLUME PRESSURE
TIME mL/hr mL/hr mL PSI
0:00:10 35.6 36.0 .10 5.8
0:00:21 35.0 34.2 .20 5.9
0:00:31 35.4 34.8 .30 5.8
0:00:42 34.6 35.1 .41 5.8
0:00:52 36.1 35.3 .51 5.8
0:01:05 28.1 34.3 .62 5.8
END OF DATA
CHANNEL 1 FLOW RATE SUMMARY
TOTAL TIME 0:01:05
TOTAL VOLUME .62mL
AVERAGE FLOW 34.3mL/hr
BACK PRESSURE 5.8PSI
NOTE:
C
This sample shows how the Log
Interval and Log Data Every
parameters in USER SETUP can
affect when data is stored in the
Datalog for a sample test.
DD
DD
D
AA
AA
A
TT
TT
T
ALAL
ALAL
AL
OGOG
OGOG
OG
DATALOG OVERVIEW
Samples of Datalogs for each test type as well as detailed
instructions for manipulating the Datalogs can be found within
each test section of this manual.