DATALOG OVERVIEW
The Datalogger is used to store data as it becomes available
in the system. There are separate Datalogs for Channel 1,
Channel 2 and the Safety Analyzer. For Channel 1 and Channel
2, the type of test determines when data is available. For Distal
Occlusion tests, data is available every second. For Flow tests,
flow rate affects when data is available. The slower the flow
rate, the more time between available data. For the Safety
Analyzer, data is available at the end of each Automatic test.
For Flow and Distal Occlusion tests, the amount of data stored
during an extended test period may fill the Datalogger. When
the Datalogger becomes full, new data is written over existing
data, beginning with the oldest record. A special function has
been incorporated into the system to restrict how often available
data is stored in the Datalog. For example, when running a 6
hour Flow test, it may only be of interest to store data every 10
seconds. To do this, the LOG INTERVAL parameter would
be set to MANUAL and the LOG DATA EVERY parameter
would be set to 00:10 (See USER SETUP for details). In
this mode, the data would be stored as it becomes available,
DD
DD
D
AA
AA
A
TT
TT
T
ALAL
ALAL
AL
OGOG
OGOG
OG
DATALOG OVERVIEW