Connectors
There are two connectors for test cables on the unit. One is for the Chassis lead and
the other is for one of two different leads used for external testing. The test cables
simply plug into the sockets. There is a release pin on the cable plug that must be
depressed to remove the cable.
Fuse
There is a fuse in the ground leg of the Analyzer Test Receptacle. This is to help
prevent damage from excess ground current. It is located on the face for ease of
replacement.
Test Receptacle
This receptacle is for the connection of the DUT. The Receptacle Rating depends on
the specific Analyzer model. SA-2001 models use a Hospital Grade North
American/USA standard NEMA 5-20R receptacle rated 20 Amps @ 125 VAC. SA-
2001-INTL models use a universal international receptacle rated 20 Amps @ 250 VAC.
SA-2001-AUS models use an Australian-specific AU1-10R receptacle rated 10 Amps @
240 VAC. An external patch cord may be necessary to connect devices utilizing
different types of plugs to the Analyzer receptacle.
Power Cord
The Power Cord, which is connected internally, provides power to both the Safety
Analyzer and the DUT through the Test Receptacle. The Power Cord varies depending
on the Analyzer model. SA-2001 models use a NEMA 5-15P plug designed to plug into
a NEMA 5-15R or 5-20R Receptacle. SA-2001-INTL models have a short IEC C20 plug
for which a country-specific adapter cable must be connected See the following
section for this information. SA-2001-AUS models use an Australian AU1-10P plug
intended for operation with Australian and New Zealand AU1-10R Receptacles.